GAD65 Antibody ELISA With Extended Reportable Range: Validation and Guidance for Neurological Practice
Abstract
OBJECTIVE: To (1) validate GAD65-ELISA detection and quantification for type 1 diabetes mellitus and autoimmune neurological diagnoses, (2) correlate ELISA results (reference range < 5 IU/mL) with established radioimmunoprecipitation assay (RIA; ≤ 0.02 nmol/L), and (3) define ELISA clinical utility and pitfalls. METHODS: Serum performance for diabetes (FDA-cleared, undiluted) was verified, and neurological laboratory-developed serum and CSF dilution protocols were validated to extend the reportable range beyond > 250 IU/mL. ELISA and RIA values were correlated, including established neurological RIA cut-offs (serum ≥ 20 nmol/L; CSF any positive). RESULTS: = 0.36). Positive agreement for RIA-defined neurological disease was 100% in serum and CSF; all serums were > 10,000 IU/mL. Clinical specificity was 97.5% in serum and 100% in CSF, exceeding reported RIA specificity. Screen results > 250 IU/mL spanned a wide range of dilution values; many were below the neurological cut-off. Most patients with paired serum/CSF positivity showed elevated GAD65 IgG indices. INTERPRETATION: GAD65 ELISA and RIA have equivalent sensitivity for autoimmune diabetes and neurological testing, with higher specificity for ELISA. A serum cut-off of 10,000 IU/mL is informative but requires clinical context, and dilution of screen > 250 IU/mL samples is essential for neurological interpretation.
Fields of Study & Subjects
Diabetes and associated disorders
Diabetes Management and Research
Pancreatic function and diabetes
Publication details
- Journal / source
- Annals of Clinical and Translational Neurology
- Publisher
- Wiley
- Publication date
- 28 March 2026
- Language
- EN