Real‐Time 3D Ultrasound Imaging with an Ultra‐Sparse, Low Power Architecture

Read full text
Authors: Colin Marcus, Md Osman Goni Nayeem, Aastha Shah, Jason Hou, S Viswanath, Maya Eusebio, David Sadat, Anantha P. Chandrakasan, Tolga Özmen, Canan Dağdeviren
Cited by: 3
References: 25

Abstract

ABSTRACT Effective resource‐constrained volumetric ultrasound imaging requires compact, low‐power systems capable of wide‐angle real‐time 3D imaging to accommodate small changes in placement by the operator. However, obtaining such images requires an excessive O(N 2 ) channel count, bulky electronics, and high power consumption. We introduce an end‐to‐end system architecture to enable high‐resolution, real‐time 3D ultrasound imaging in a portable form factor. We present: a convolutional optimally distributed array (CODA) geometry that drastically reduces the number of elements (from 1024 to 128), a novel chirped data acquisition (cDAQ) architecture that enhances imaging depth while operating with a 25.3 dB lower transmit amplitude than a pulsed system, and an associated new signal processing methodology. We experimentally demonstrate our system's ability to perform deep (> 11 cm), high axial resolution (< 600 µm), and wide‐angle (57°) imaging, while simultaneously reducing power consumption (29.6x reduction) and drive voltage (18 V). We validated our system in vitro and further performed in vivo human trials, demonstrating the ability to detect both tumors and cysts in breast tissue. This new architectural approach will unlock a new class of medical devices with enhanced diagnostic and long‐term monitoring capabilities and open up future wearable designs of real‐time 3D ultrasound systems.

Fields of Study & Subjects

Ultrasound Imaging and Elastography Photoacoustic and Ultrasonic Imaging Ultrasound and Hyperthermia Applications

Publication details

Journal / source
Advanced Healthcare Materials
Publisher
Wiley
Volume
15
Issue
16
Pages
e05310–e05310
Publication date
29 January 2026
Language
EN