Repositioning of a Protek Duo Temporary Right Ventricular Assist Device Using Balloon‐Assisted Tracking Technique

Read full text
Authors: Kevin A. Friede, Joseph S. Rossi, Michael Halista
Cited by: 0

Abstract

Right ventricular failure after left ventricular assist device implantation is associated with significant morbidity and mortality. The use of percutaneous right ventricular assist devices for treatment in this clinical situation is increasing, but these devices are prone to migration with limited options for revision or repositioning. We describe a case of a patient whose device was repositioned safely using a balloon-assisted tracking technique with a peripheral angioplasty balloon, analogous to the technique developed to advance catheters past radial artery stenoses in coronary interventions.

Fields of Study & Subjects

Mechanical Circulatory Support Devices Cardiac pacing and defibrillation studies Congenital Heart Disease Studies

Publication details

Journal / source
Catheterization and Cardiovascular Interventions
Volume
107
Issue
5
Pages
1503–1505
Publication date
8 February 2026
Language
EN