Novel Impedance‐Guided Contact Mapping Technique for the Circular Multielectrode Pulsed‐Field Ablation Catheter
Abstract
Impedance-guided PulseSelect contact mapping (i-Pulse) adapts EnSite X Contact Index for a circular multielectrode PFA catheter, enabling real-time visualization of electrode-tissue contact without fluoroscopy. With ICE guidance, Δimpedance > 5% identified adequate contact, standardized tagging, and guided applications in three atrial fibrillation cases, achieving acute pulmonary vein isolation without complications.
Fields of Study & Subjects
Cardiac Arrhythmias and Treatments
Vacuum and Plasma Arcs
Diamond and Carbon-based Materials Research
Publication details
- Journal / source
- Clinical Case Reports
- Publisher
- Wiley
- Volume
- 14
- Issue
- 3
- Pages
- e72142–e72142
- Publication date
- 24 February 2026
- Language
- EN