The MAIIA EPO Purification Gel Kit, 7D3—An Alternative Purification Kit for Screening and Confirmation Procedures. Validation for Serum and Plasma Matrices

View at publisher
Authors: Jana Hempel, Dirk Schwenke, S Reimann, S C Voss
Cited by: 0
References: 13

Abstract

In this study, the performance of the MAIIA EPO purification kit 7D3 was evaluated for human serum and plasma matrices. The kit was validated for selectivity in comparison to the MAIIA 3F6 kit, reliability at minimum required performance levels (MRPLs), limit of detection (LOD), carryover and recovery to meet compliance with the World Anti-Doping Agency criteria of the technical document 2024 (TD2024EPO). In addition, the influence of haemolysis on plasma samples was evaluated. In total, 30 serum samples and 30 plasma samples were tested for selectivity; 10 serum and 10 plasma samples, for reliability at MRPL and recovery; and six serum and six plasma samples, for LOD. The influence of haemolysis was evaluated by creating mixtures of 0%, 25%, 50%, 75% and 100% of a hemolytic sample and a blank sample, and the same sample spiked at the MRPL level. This study validates that the 7D3 kit meets all WADA TD2024EPO criteria, providing a reliable alternative for the purification of erythropoietin receptor agonists from human serum and plasma matrices in both screening and confirmation procedures in routine antidoping analysis.

Fields of Study & Subjects

Doping in Sports Pharmacological Effects and Assays Biosimilars and Bioanalytical Methods

Publication details

Journal / source
Drug Testing and Analysis
Volume
18
Issue
4
Pages
552–560
Publication date
24 February 2026
Language
EN