BACKGROUND: Artefacts in contrast-enhanced mammography (CEM) can compromise diagnostic image quality. These artefacts may obscure clinically relevant findings or mimic pathological features, potentially leading to misinterpretation. PURPOSE: A virtual imaging trial technique was used to quantify the impact of breast-in-breast and motion artefacts on lesion contrast and signal-difference-to-noise ratio (SDNR) in CEM. METHODS: Computational breast phantoms representing four breast density categories were generated, and 5 mm diameter non-spiculated masses were embedded centrally and peripherally to study the influence of density and the breast-in-breast artefact. Motion artefacts were simulated using two techniques: horizontal and vertical translation of the entire phantom between the low-energy (LE) and high-energy (HE) acquisitions, and simulation of partial paddle release between the two exposures. A Monte Carlo (MC) method was used to model the LE and HE images generated by a Siemens MAMMOMAT B.brilliant system. An offline algorithm with motion correction generated the recombined images. Lesion contrast and SDNR were evaluated for simulated masses with clinically relevant iodine concentrations derived from patient data and QC phantoms. RESULTS: The breast-in-breast artefact causes contrast and SDNR reductions of up to 76% for peripheral lesions. There was a slight increase in these parameters as breast density increased, due to the fixed weighting factor used by the recombination algorithm. Simulated rigid and non-rigid motion reduced SDNR by up to 60% before motion correction. Activation of the motion correction algorithm preserved SDNR for small in-plane displacements, although some residual artefacts remained for non-rigid motion. CONCLUSIONS: This study highlights the challenges posed by non-uniform breast thickness at the borders and emphasizes the importance of motion correction. Virtual imaging trials provide a powerful, controlled method to study these effects, offering insights that are challenging to obtain from clinical or physical phantom studies alone.

Fields of Study & Subjects

Digital Radiography and Breast Imaging Advanced X-ray Imaging Techniques Advanced X-ray and CT Imaging

Publication details

Journal / source
Medical Physics
Publisher
Wiley
Volume
53
Issue
2
Pages
e70326–e70326
Publication date
1 February 2026
Language
EN