Chemical Shift Separated and Compensated Ultra‐Short Echo‐Time Imaging

Read full text
Authors: Martin Krämer, Lumeng Cui, Jürgen Rainer Reichenbach, Stefan Sommer
Cited by: 1
References: 27

Abstract

PURPOSE: To develop a modified binomial excitation scheme for ultra-short echo-time (UTE) imaging that allows for the separation of water and fat signals and the correction of chemical shift artifacts. METHODS: Theoretical derivation and numerical Bloch simulations were performed for a modified water-excitation binomial pulse scheme to calculate fat and water signals as well as a chemical shift corrected UTE image from two acquisitions with these configurations. Chemical shift correction was applied at the readout level by multiplying the fat signal with a corresponding linear phase term. Measurements with the proposed binomial pulse scheme were performed at 3 T and 7 T. RESULTS: Imaging results at 7 T and 3 T showed robust fat and water separation and images free of chemical shift-induced blurring after applying the proposed method. Compared to standard UTE, the corrected UTE images exhibited improved tissue boundary delineation and improved visibility of fine details. Bloch simulations demonstrated that deviations of the phase differences between water and fat from the analytical solution were below 1% for typical UTE acquisition parameters. CONCLUSION: The proposed binomial excitation scheme and post-processing effectively separate water and fat signals and correct for chemical shift artifacts in UTE imaging, providing improved image quality, especially at high field strengths.

Fields of Study & Subjects

Advanced MRI Techniques and Applications Electron Spin Resonance Studies Integrated Circuits and Semiconductor Failure Analysis

Publication details

Journal / source
Magnetic Resonance in Medicine
Publisher
Wiley
Volume
96
Issue
1
Pages
121–133
Publication date
27 February 2026
Language
EN