Abstract

PURPOSE: Phase singularities are a common problem in image reconstruction with auto-calibrated sensitivities due to an inherent ambiguity of the estimation problem. The purpose of this work is to develop a method for detecting and correcting phase poles in non-linear inverse (NLINV) reconstruction of MR images and coil sensitivity maps. METHODS: Phase poles are detected in individual coil sensitivity maps by computing the curl in each pixel. A weighted average of the curl in each coil is computed to detect phase poles. Phase pole detection and correction is then integrated into the iteratively regularized Gauss-Newton method of the NLINV algorithm. In addition, we demonstrate the algorithm can also remove phase poles in ESPIRiT coil sensitivity maps. Phase pole correction is evaluated for reconstruction of accelerated Cartesian MPRAGE data of the brain and interactive radial real-time MRI of the human heart. RESULTS: ) auto-calibration (AC) regions. CONCLUSION: With the proposed method, phase poles can be reliably removed in reconstructions and coil sensitivities.

Fields of Study & Subjects

Advanced MRI Techniques and Applications Functional Brain Connectivity Studies NMR spectroscopy and applications

Publication details

Journal / source
Magnetic Resonance in Medicine
Publisher
Wiley
Volume
96
Issue
1
Pages
134–145
Publication date
12 March 2026