BART Streams: Real‐Time Reconstruction Using a Modular Framework for Pipeline Processing
Abstract
PURPOSE: To create modular solutions for interactive real-time MRI using reconstruction algorithms implemented in BART. METHODS: A new protocol for streaming of multidimensional arrays is presented and integrated into BART. The new functionality is demonstrated using examples for cardiac interactive real-time MRI based on radial FLASH, where iterative reconstruction is combined with advanced features such as dynamic coil compression and gradient-delay correction. We analyze the latency of the reconstruction and measure end-to-end latency of the full imaging process. RESULTS: Reconstruction pipelines with iterative reconstruction and advanced functionality were built in a modular way using scripting. Latency measurements demonstrate latency sufficient for interactive real-time MRI, on the order of 30 ms for BART processing and network transfer time, or 200 ms for end-to-end latency including acquisition, vendor processing, and display. CONCLUSION: With the new streaming capabilities, real-time reconstruction pipelines can be assembled using BART in a flexible way, enabling rapid prototyping of advanced applications such as interactive real-time MRI.
Fields of Study & Subjects
Advanced MRI Techniques and Applications
Medical Imaging Techniques and Applications
Functional Brain Connectivity Studies
Publication details
- Journal / source
- Magnetic Resonance in Medicine
- Publisher
- Wiley
- Volume
- 96
- Issue
- 4
- Pages
- 1534–1544
- Publication date
- 4 June 2026