Attenuation-based ultra-low-dose lung computed tomography at 0.1 mSv to 0.3 mSv effective dose in children
Abstract
BACKGROUND: Radiation dose reduction is essential in paediatric lung computed tomography (CT). Advances in energy-integrating detector CT and deep-learning reconstruction may enable ultra-low-dose imaging comparable to photon-counting CT. OBJECTIVE: To evaluate the radiation dose and performance of an ultra-low-dose lung CT protocol using a wide-detector energy-integrating CT system in paediatric patients, focusing on effective radiation dose and diagnostic image quality. MATERIALS AND METHODS: A total of 277 low-dose lung CT scans from 106 paediatric patients (age range, 113 days to 17.85 years) were retrospectively analysed. All scans were acquired in axial mode using a 256-slice-multidetector CT scanner with deep learning image reconstruction and attenuation-based Auto Prescription. Radiation dose parameters, including volume CT dose index, dose-length product, size-specific dose estimate, and effective dose, were calculated. Signal-to-noise ratio and contrast-to-noise ratio were assessed in standardised anatomical regions. Patients were stratified by age, and statistical analysis was conducted to evaluate dose trends and image quality metrics. RESULTS: There were significant differences between all age groups for all dose parameters (Kruskal-Wallis test, P
Fields of Study & Subjects
Advanced X-ray and CT Imaging
Radiation Dose and Imaging
Digital Radiography and Breast Imaging
Publication details
- Journal / source
- Pediatric Radiology
- Publisher
- Springer Science and Business Media LLC
- Volume
- 56
- Issue
- 3
- Pages
- 536–544
- Publication date
- 19 January 2026
- Language
- EN