Cardiac MR function analysis with DL-based super resolution reconstruction: application in the clinical setting

Read full text
Authors: Franziska Adomat, Christof Schaub, Tobias Hoh, Xenia Fischer, Roman Guggenberger, R. H. Manka, Matthias Eberhard, Lucas Weber
Cited by: 1
References: 36

Abstract

To assess differences in volumetry, image quality and acquisition time between balanced steady-state free precession cine sequences acquired using (a) a standardized sensitivity encoding (SENSE) approach and (b) deep learning-based super-resolution reconstruction based on high-resolution images acquired with compressed sensitivity encoding (C-SENSE). We retrospectively evaluated 31 consecutive patients (mean age 61.2 ± 13.1 years, 26% female (8/31) and 74% male (23/31)) undergoing cardiac magnetic resonance imaging (MRI) examinations to assess for the presence of ischemic and non-ischemic cardiomyopathies. Cine images were acquired using a 1.5T Philips Ingenia MRI scanner, with classic parallel imaging (SENSE) and compressed sensing (C-SENSE) accelerated acquisition techniques (R = 2 and R = 4, respectively). C-SENSE datasets were reconstructed using a deep learning-based denoising and super-resolution algorithm to enhance image resolution and quality (CS-SR). To evaluate cardiac function, manual left ventricular (LV) segmentation and volumetric analysis were performed on both datasets by two readers, who were blinded to the clinical data. Image quality was rated independently by three readers using Likert scales. Correlation between SENSE and CS-SR datasets with respect to LV volumetry was high (r = 0.98-1.00), with no significant differences found for end-diastolic volume (mean difference 0.04 ml, limits of agreement (LoA) -11.19 to 11.26 ml; p = 0.970) or end-systolic volume (mean difference 1.60 ml, LoA - 7.48 to 10.68 ml; p = 0.064). Overall subjective image quality was comparable (p = 0.061), with CS-SR offering better image sharpness at the cost of increased artifacts (p < 0.001 respectively). Image acquisition time was significantly accelerated with C-SENSE acquisition (SENSE: 411.1 ± 47.7 s, C-SENSE: 165.6 ± 21.5 s; p < 0.001). CS-SR shows promise in streamlining routine cardiac imaging by significantly shortening acquisition times, without impairing LV volumetric analysis, while preserving overall image quality and resolution.

Fields of Study & Subjects

Cardiac Imaging and Diagnostics Advanced MRI Techniques and Applications Cardiovascular Function and Risk Factors

Publication details

Journal / source
The International Journal of Cardiovascular Imaging
Publisher
Springer Science and Business Media LLC
Volume
42
Issue
5
Pages
953–963
Publication date
9 February 2026
Language
EN