Intercomparison exercise of easy-to-measure and non-volatile difficult-to-measure analysis in homogenised high activity steel

Read full text
Authors: Anumaija Leskinen, Simon Jerome, Nancy Nazem Wanna, Céline Gautier, Veronika Meriläinen, Jussi Leporanta, Antti Räty, Lesley Adriaensen, Ingrid Guens
Cited by: 1
References: 32

Abstract

Abstract Intercomparison exercises (IE) are a way for laboratories to test their analytical performance. In case of difficult-to-measure (DTM) radionuclide analysis, participation in IEs are important due to the lack of reference materials. This paper reports the results from an IE focusing on non-volatile DTM analysis in homogenised high activity steel. The IE was carried out according to the ISO 13528 standard, in which the performances are assessed using z score. The z scores were generally acceptable. The experimental results were compared with activation calculation results showing wide range of compatibility.

Fields of Study & Subjects

Scientific Measurement and Uncertainty Evaluation Radioactive Decay and Measurement Techniques Pesticide Residue Analysis and Safety

Publication details

Journal / source
Journal of Radioanalytical and Nuclear Chemistry
Publisher
Springer Science and Business Media LLC
Volume
335
Issue
1
Pages
351–374
Publication date
1 January 2026
Language