Public Defibrillator Accessibility in Slovakia: A Descriptive Study

Read full text
Authors: Adrián Fabiny
Cited by: 0
References: 20

Abstract

To analyse the Slovak National AED Registry for critical gaps in device accessibility, availability, and specifications based on 2025 ERC Guidelines. Out-of-Hospital Cardiac Arrest (OHCA) is a time-critical challenge where survival depends on Public Access Defibrillation (PAD). Slovakia has a national AED registry, but its compliance with international recommendations is unknown. A descriptive, cross-sectional analysis of all registered AEDs (N = 4484) as of August 2025 was conducted. We analysed device type, paediatric capability, 24/7 availability, location, and physical barriers (mechanical locks). The registry included 4484 AEDs. While 71.1% (N = 3187) were available 24/7, only 20.2% (N = 904) had a paediatric mode. Merely 4.9% (N = 221) had both paediatric capability and 24/7 availability. Furthermore, 7.0% (N = 314) were secured with a mechanical lock. “Barrier-free access” data (90.3%) was misleading, not referring to device placement. The Slovak AED network has critical deficiencies contradicting ERC 2025 recommendations. The extremely low 24/7 availability of paediatric-capable AEDs (4.9%) and the presence of mechanical locks (7.0%) limit PAD effectiveness. Urgent technical regulations are necessary.

Fields of Study & Subjects

Cardiac Arrest and Resuscitation Cardiac pacing and defibrillation studies Mechanical Circulatory Support Devices

Publication details

Journal / source
Bratislavské lekárske listy/Bratislava medical journal
Volume
127
Issue
3
Pages
1302–1308
Publication date
7 January 2026
Language
EN