Assessing the imaging calibration factor for quantitative SPECT/CT of actinium-225

Read full text
Authors: Monika Tulik, Jakub Krzemiński, Dariusz Pawlak, Frank Bruchertseifer, Livia Chilug, Jakub Marek Baran, Jolanta Kunikowska
Cited by: 0
References: 34

Abstract

Objective The main aim of this study was to evaluate the effect of quantitative imaging protocol parameters on the imaging calibration factor (ICF) value for 225 Ac. Approach All measurements were conducted with two measurement geometries: point source (activity of 14.9 MBq) and cylinder filled uniformly with radionuclide (total injected activity of 41.4 MBq). The data were acquired as follows: planar acquisition of a point source (duration 20 min) and tomographic (SPECT/CT) acquisition of a cylindrical phantom (total acquisition time 32 min or 56 min, matrix size 128×128). Nine different combinations of primary and scatter energy windows for 225 Ac quantitative imaging were studied. All acquired data were re-analyzed for a 64×64 matrix size. The aforementioned tomographic experiment was repeated when the total activity in the cylindrical phantom dropped to 4.2 MBq, and for selected combinations of energy windows. ICFs for planar images were determined by taking the total number of counts in the whole field of view of the camera. ICFs for SPECT/CT images were determined with four different methods of segmentation. Additionally, the homogeneity of the reconstructed SPECT/CT images was evaluated. Main results and significance The ICF values obtained using planar/tomographic geometries and different configurations of acquisition settings were presented. Based on the results obtained, for the 225 Ac quantitative acquisition protocol, it could be recommended to define two main energy windows (for 213 Bi and 221 Fr, daughter nuclei of 225 Ac) with accompanying lower and upper energy windows for TEW scattering correction. The benefits of a 64×64 matrix should also be considered.

Fields of Study & Subjects

Medical Imaging Techniques and Applications Advanced X-ray and CT Imaging Radioactive Decay and Measurement Techniques

Publication details

Journal / source
Applied Radiation and Isotopes
Publisher
Elsevier BV
Volume
236
Pages
112773–112773
Publication date
23 June 2026
Language
EN