Thinner is not always better: Focal in-stent restenosis with stent deformation at coronary bend following ultrathin-strut drug-eluting stent implantation

Read full text
Authors: Asharam Panda, Sumit Kumar, SS Mishra
Cited by: 0
References: 4

Abstract

We describe a 64-year-old male who underwent percutaneous coronary intervention of an angulated right coronary artery lesion using overlapping ultrathin-strut everolimus-eluting stents. Although the initial angiographic result was optimal, the patient developed recurrent angina at six months, and repeat angiography revealed focal in-stent restenosis localised to the coronary bends with fluoroscopic evidence of stent deformation. The overlap segment was relatively spared. Repeat intervention with a conventional thin-strut DES resulted in excellent clinical and angiographic outcomes at one-year follow-up. This case underscores the importance of unfavourable outcome of ultrathin strut DES at coronary bends, possibly secondary to poor radial strength leading to stent deformation and in-stent restenosis.

Fields of Study & Subjects

Coronary Interventions and Diagnostics Cardiac Imaging and Diagnostics Acute Myocardial Infarction Research

Publication details

Journal / source
Cardiovascular Revascularization Medicine Interesting Cases
Volume
11
Pages
100142–100142
Publication date
10 February 2026
Language
EN