Failure analysis of retrieved stainless steel Exeter hip implants: A fractographic and corrosion perspective

Read full text
Authors: Saman Nikpour, Vahid Dehnavi, Brent A. Lanting, Matthew G. Teeter, Fuad Khoshnaw, Yolanda Susanne Hedberg
Cited by: 0
References: 40

Abstract

• Fatigue was the dominating failure mechanism for the three cemented REX 734 stems. • Cracks consistently initiated at the anterolateral region near the mid-length of the stem. • Coarse Nb-enriched phases likely assisted crack initiation and propagation. • Insufficient cement fixation likely enabled fluid ingress and corrosion fatigue. • Patient weight and BMI influenced crack propagation and the area of final fracture. In this study, three fractured stainless steel Exeter (REX 734) hip joint implants were analyzed to identify the underlying failure mechanisms. These cemented hip stems with a polished surface finish failed earlier than their expected service life, indicated by sudden pain in the patients, necessitating revision surgeries. Optical and scanning electron microscopy were used to analyze the explanted hip stem fracture surfaces. The fractograms exhibited three distinctive zones; (1) crack initiation at the anterolateral part of the stem, (2) crack propagation zone with beach marks and striations, and (3) final fracture zone. Fatigue was the dominant failure mechanism. The anterolateral surface of the implants close to the fracture initiation site showed discoloration and surface alteration, possibly due to corrosion, and the presence of extrusions and intrusions formed by slip band accumulation, causing stress concentration and crack initiation. Also, elongated and sharp-edged phases enriched in niobium were found at the crack initiation site and throughout the microstructure. Medical radiographs indicated a lack of cement fixation at the proximal part of the stems, causing implant instability and ingress of body fluids, which potentially led to corrosion reactions. The crack propagation zone decreased in relative size with higher patient weight and BMI. Overall, this study suggests the possibility of corrosion contributions and a role of niobium-rich precipitates in the crack initiation and fatigue failure.

Fields of Study & Subjects

Orthopaedic implants and arthroplasty Mechanical Failure Analysis and Simulation Titanium Alloys Microstructure and Properties

Publication details

Journal / source
Engineering Failure Analysis
Volume
190
Pages
110718–110718
Publication date
7 March 2026
Language
EN