Simple programming, significant impact: Pacemaker Sleep Function extends device longevity

Read full text
Authors: Kazuma CHIKURA, Daisuke Kawano, Hiroshi Taniguchi, Takayuki Miyake, Masami Sakurada, Hitoshi Mori
Cited by: 1
References: 8

Abstract

Pacemaker (PM) battery longevity is a clinically important issue, particularly in elderly patients at higher risk for device replacement-related complications. Sleep function, a sleep-specific rate reduction feature, lowers the programmed lower rate limit (LRL) during the sleep period and thereby suppresses unnecessary pacing. We present a unique case of a 90-year-old man in whom activation of the sleep-specific rate reduction feature significantly reduced excessive atrial pacing, resulting in an estimated one-year extension of battery longevity. Sleep-specific rate reduction features during sleep periods could be a simple and effective strategy to prolong PM battery longevity.

Fields of Study & Subjects

Sleep and Work-Related Fatigue Obstructive Sleep Apnea Research Restless Legs Syndrome Research

Publication details

Journal / source
HeartRhythm Case Reports
Publisher
Elsevier BV
Volume
12
Issue
6
Pages
639–642
Publication date
1 April 2026
Language