Successful Extraction of Entrapped CIED Lead After TTVR

Read full text
Authors: Johannes Kirchner, Barbara Rubinić, Muhammed Gerçek, Volker Rudolph, Thomas Eitz, Thomas Fink, Christian Sohns, Philipp Sommer, Guram Imnadze, Tanja K. Rudolph
Cited by: 1
References: 4

Abstract

BACKGROUND: Transcatheter tricuspid valve replacement (TTVR) has emerged as a therapeutic option for patients with symptomatic severe tricuspid regurgitation. Right ventricular (RV) leads from pre-existing cardiac implantable electronic devices are entrapped during TTVR device implantation. CASE SUMMARY: An 83-year-old woman with a dual-chamber cardiac implantable electronic device underwent TTVR in 2022. The procedure resulted in entrapment of the RV lead. After TTVR, the increase in lead impedance raised concern for lead failure, prompting consideration of a leadless pacemaker. During this procedure, the entrapped lead was extracted without causing malposition of the TTVR device. DISCUSSION: Extraction of entrapped leads after TTVR may be feasible in experienced centers after cautious deliberation. Nevertheless, pre-emptive lead extraction should be considered to minimize the risk of unsuccessful extraction or serious complications. TAKE-HOME MESSAGE: Extraction of entrapped RV leads without causing TTVR device malposition may be feasible, and leadless pacemakers can be considered a technically viable alternative for RV pacing in TTVR.

Fields of Study & Subjects

Cardiac pacing and defibrillation studies Electrostatic Discharge in Electronics Integrated Circuits and Semiconductor Failure Analysis

Publication details

Journal / source
JACC Case Reports
Publisher
Elsevier BV
Volume
31
Issue
9
Pages
106583–106583
Publication date
21 January 2026
Language
EN