A hyperelastic-plastic damage model for puncture analysis of tympanic membrane using finite-element method

Read full text
Authors: Hossein Mohammadi, Nima Maftoon
Cited by: 1
References: 67

Abstract

The puncture and damage behaviour of the tympanic membrane (TM) during needle insertion, a common procedure for accessing the middle and inner ear, is critical for advancing treatments for hearing and balance disorders. This study presents a novel 3D finite-element (FE) model of TM puncture, capturing both the geometric complexity of the needle and the middle ear and the nonlinear material properties of the TM, including its high-strain deformation, damage accumulation, and crack propagation. By calibrating damage-related mechanical properties, including fracture strain and yield stress, the presented model accurately replicated experimental puncture forces (2.6% error) and predicted a maximum von Mises stress of 1.0 MPa at the puncture instance. The model also enabled analysis of the effects of insertion location, needle inclination, needle tip geometry, and TM thickness on puncture force. Additionally, it allowed for the evaluation of both axial and lateral forces, offering insights for the development of haptic devices, surgical robots, and virtual-reality medical training systems.

Fields of Study & Subjects

Ear Surgery and Otitis Media Soft Robotics and Applications Speech and Audio Processing

Publication details

Journal / source
Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials
Publisher
Elsevier BV
Volume
180
Pages
107461–107461
Publication date
8 May 2026
Language
EN