A high-resolution ionization chamber for measuring Radon in ambient air

Read full text
Authors: Ralf Nötzel, Kerstin Weinberg
Cited by: 1
References: 7

Abstract

Radon is a naturally occurring radioactive gas that contributes significantly to human radiation exposure and must be controlled to avoid concentrations harmful to health. The paper presents an impulse-proportional ionization chamber that is suitable for the direct measurement of low radon concentrations in ambient air and achieves spectroscopic resolutions of 2 %. This accuracy can be achieved through two novel principles. Firstly, the double-cylindrical, coaxial design of the IC allows for efficient, nearly complete detection of α -radiation. Secondly, a customized measurement method for spectroscopic evaluation was developed to discriminate between the proportions of 222 Rn , 218 Po , and 214 Po and extract their concentrations. Particular attention was paid to the energy resolution of the detection system by suppressing the effects of acoustic and vibration noise on the detector’s operation. The high spectral resolution of the developed ionization chambers, with working volumes of 7.7 and 8.8 l, enables measurements with uncertainties of less than 5% at 15-minute measurement times in ambient air with 50 Bq/m 3 radon activity.

Fields of Study & Subjects

Radioactivity and Radon Measurements Radiation Detection and Scintillator Technologies Radioactive contamination and transfer

Publication details

Journal / source
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Publisher
Elsevier BV
Volume
1090
Pages
171617–171617
Publication date
7 May 2026
Language
EN