Development and application of in-vivo dose and time-resolved measurements for clinical application of ultra-high dose rate radiotherapy

Read full text Summarise with VIRAT
Authors: Riccardo Dal Bello, S. Psoroulas, Dominik Flückiger, Jerome Krayenbuehl, Arvid Kemper, Rafael Kranzer, Benjamin Côté, Jens von der Grün, Panagiotis Balermpas, Matthias Guckenberger, Stephanie Tanadini‐Lang
Cited by: 0
References: 42

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

Background and purpose: dose monitoring in a clinical setting. Materials and methods: Four detector systems were evaluated for patient treatments with UHDR at a modified C-arm linear accelerator (linac) (9 MeV, dose per pulse 1.08 Gy, average dose rate 216 Gy/s): two passive systems - optically stimulated luminescence detectors (OSLDs) and radiochromic films - and two active systems - a scintillator and a diamond detector. The detectors were positioned on a dedicated 3D-printed mount or the patient's skin to measure integral dose and time-resolved parameters. OSLDs were calibrated against a reference ion chamber and validated under varying conditions. Scintillator and diamond detectors recorded pulse-to-pulse and intra-pulse stability, respectively. Results: OSLDs provided reproducible dose reporting within ±5%, with precision improved through averaging multiple chips. The pulse-to-pulse stability measured with the scintillator was in agreement with the diamond detector measurements within ±1%. The diamond detector resolved intra-pulse stability and identified instabilities in case of poor beam tuning. Conclusions: The combined use of OSLDs, scintillator, diamond, and film detectors enabled robust and time-resolved recording of UHDR RT in a clinical trial setting. While resource-intensive, this approach ensured redundancy and precision, providing a feasible strategy for early clinical adoption.

Fields of Study & Subjects

Advanced Radiotherapy Techniques Radiation Therapy and Dosimetry Radiation Dose and Imaging

Publication details

Journal / source
Physics and Imaging in Radiation Oncology
Volume
38
Pages
100959–100959
Publication date
1 March 2026
Language
EN