Powering smart orthopedic implants using acoustic waves: An ex vivo study

Read full text
Authors: Olivier Freychet, N. Garraud, Martial Defoort, Samuel Guigo, Valérie Burdin, Skandar Basrour, H. Letissier, G. Dardenne, P. Gasnier
Cited by: 1
References: 37

Abstract

This work experimentally investigates the feasibility of acoustic power transfer for battery-free full-metal smart orthopedic implants. Two receiver placements were evaluated on both knees of a human cadaver: a stem placement, where the acoustic path includes cancellous bone, and a tray placement, where the acoustic path avoids bone but integration constraints are more stringent. Receivers with resonance frequencies between 250 kHz and 2 MHz were designed using finite element simulations, analytical modelling, and phantom experiments. Measurements show that increasing frequency significantly reduces transmitted power due to higher attenuation in cancellous bone. It was shown that up to 29 mW and 20 mW of power could be transferred with the stem and tray placement, respectively, with an acoustic intensity of 720 mW/cm 2 and mechanical index below 1.9, hence respecting exposure limits. Volumetric power densities of 91 mW/cm 3 and 1400 mW/cm 3 were reached with the stem and tray placement, respectively, which are comparable or higher than those reached with electromagnetic techniques (from 1.16 to 194 mW/cm 3 ) while compliance with electromagnetic exposure limits is not always clearly addressed in the studies. This work demonstrates the feasibility of milliwatt-level acoustic powering of smart orthopedic implants while highlighting challenges related to transmitter alignment and narrow operating bandwidth.

Fields of Study & Subjects

Wireless Power Transfer Systems Wireless Body Area Networks Muscle activation and electromyography studies

Publication details

Journal / source
Ultrasonics
Publisher
Elsevier BV
Volume
168
Pages
108231–108231
Publication date
10 July 2026
Language
EN