Development and application of a structured preclinical protocol for MRI-guided endomyocardial biopsy: a multimodal imaging approach using a hybrid vessel phantom and porcine in vivo models

Read full text
Authors: Denis Gholami Bajestani, C. Martin Reich, Jan Birkigt, Christina Mulik, Grit Gesine Ruth Hiller, H Busse, Timm Denecke, Johannes Kauffold, Haukur Sigmarsson, Thies H. Jochimsen, Bernhard Sattler, Osama Sabri
+2 more Karsten Lenk, Andreas Melzer
Cited by: 1
References: 17

Abstract

BACKGROUND: MRI-guided cardiovascular procedures provide high-resolution, radiation-free imaging, but clinical translation is limited due to the lack of MR Conditional medical devices. This study presents a structured preclinical evaluation protocol for MRI-guided endomyocardial biopsy (EMB) interventions, demonstrating its feasibility for both instrument performance and procedural workflow. METHODS: An MR-compatible prototype EMB forceps was evaluated using the three-stage preclinical protocol. First, device visibility was assessed using fluoroscopy and two MRI systems (Magnetom Sola 1.5T and a Biograph 3.0T, both Siemens Healthineers). Tissue samples retrieved from the left ventricle of a fresh porcine heart using MR-compatible (n = 5) and standard Cordis biopsy forceps (n = 5) were subjected to histological analysis. In the second stage, the device and workflow were tested using a developed hybrid vessel phantom, combining a human CT-based silicone aortic model and a Thiel-embalmed porcine heart. Handling performance was assessed by experienced cardiac interventionists (n = 4). Finally, in vivo testing was performed in a porcine model (n = 4), following ethics committee approval. RESULTS: Visibility of the EMB forceps was confirmed in all MRI systems and fluoroscopy. Handling was rated medium or higher by all operators. No significant difference in histological tissue quality and tissue sampling times (p > 0.05) was observed between MR-compatible and standard EMB forceps. In vivo, five tissue samples of equivalent quality were successfully harvested from six attempts, with the MR-compatible EMB forceps exhibiting smaller MRI artifact sizes in vivo compared to in vitro measurements. CONCLUSIONS: The developed preclinical testing protocol effectively evaluated key device characteristics (reliable visibility, handling, and accurate tissue sampling) of the MR-compatible EMB forceps, providing valuable insights in the early stages of in vivo trials.

Fields of Study & Subjects

Cardiac Imaging and Diagnostics Cardiomyopathy and Myosin Studies Advanced MRI Techniques and Applications

Publication details

Journal / source
Zeitschrift für Medizinische Physik
Publisher
Elsevier BV
Publication date
1 January 2026
Language
EN