The “five-second” rule for dropped food: does it apply to dropped medical objects in the operating room? A randomized study of disinfection approaches for contaminated arthroplasty implants
Summary
No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.
Summarise with VIRAT Free to try with your Science Catalogs account.Abstract
Polyethylene liners dropped onto operating room floors rapidly acquire bacterial contamination, including clinically important pathogens. In a randomized bench study, sterile chlorhexidine-alcohol and povidone-iodine immersion significantly reduced bioburden compared with ethanol or no intervention. When replacement is not feasible, chlorhexidine or iodine is preferable.