Tandem Mobility Measurements of Ionic Liquid Cluster Volatilities
Abstract
We illustrate the potential of a pair of steady-state ion mobility filters run in series, similar to a triple quadrupole mass spectrometer, not only to operate under atmospheric pressure but also to determine reaction kinetics at controlled pressure and temperature. Singly charged cluster cations of the ionic liquid 1-ethyl-3-methyl-imidazolium tris(pentafluoroethyl)trifluorophosphate (EMI-FAP) produced by a bipolar electrospray source are isolated steadily at atmospheric pressure by a first ion mobility spectrometer (IMS). A second IMS then determines the composition of evaporative decay products: (EMI-FAP) n −1 EMI + →(EMI-FAP) n −2 EMI + +EMI-FAP (2 ≤ n ≤ 30) after passing through a heated parabolic flow reactor tube at controlled temperatures T (23–40 °C). Measured product/parent abundance ratios are converted into reaction constants k n via a previously described theory. At n < 11, evaporation rates k n exhibit idiosyncrasies characteristic of various magic numbers, some of which are also reflected in a cluster compactness parameter inferred from the measured mobility cross section and mass. An approximately continuous variation of volatility with n is found for 11 ≤ n ≤ 14, and for n ≥ 19, with a region of low volatility in between. This especially stable domain of four clusters (15 ≤ n ≤ 18) is reached on the right ( n = 18) with a discontinuous (4-fold) volatility decrease, suggesting a transition to a liquid-like structure. On the left it merges continuously (though with a steep slope) with the trend at smaller sizes (11 ≤ n ≤ 14).
Fields of Study & Subjects
Mass Spectrometry Techniques and Applications
Ionic liquids properties and applications
Electrohydrodynamics and Fluid Dynamics
Publication details
- Journal / source
- Analytical Chemistry
- Volume
- 98
- Issue
- 7
- Pages
- 5284–5291
- Publication date
- 10 February 2026
- Language
- EN