Wafer-Scale Fabrication of Miniaturized Graphene-Based Ion-Selective Field-Effect Transistor Chips for Clinical Plasma Test

View at publisher
Authors: Qi Shu, Hongru Ma, Lvkun Yang, Xiaoqing Qi, Kaicong Liu, Liuna Luo, Wenjia Xie, Siqi Li, Kangxin Hou, Zijian Shi, Fuxiang Song, Donglin Ma
+2 more Gangshi Wang, Teng Gao
Cited by: 0
References: 47

Abstract

Rapid and reliable point-of-care plasma electrolyte analysis is critical for patient management under extensive care, driving demand for miniaturized sensors capable of ultralow sample consumption and multiplexed detection. We present a wafer-scale fabricated ion-selective graphene field-effect transistor (ISGFET) array that overcomes limitations of conventional ion-selective electrodes through a 2 μm × 5 μm sensing area and monolithic integration of K + /Na + /Ca 2+ selective membranes. The ISGFET sensor exhibited high sensitivity (52.43 mV/decade for K +, 53.62 mV/decade for Na +, and 28.72 mV/decade for Ca 2+ ) and selectivity in ion detection, as well as notable stability under physiological conditions and long-term reusability for over a month. The ISGFET-based sensing platform for the first time demonstrated its 100% diagnostic accuracy for potassium measurement in human plasma with an error of 4.37% (0.17 mM), when consuming less (

Fields of Study & Subjects

Analytical Chemistry and Sensors Graphene research and applications Advanced Sensor and Energy Harvesting Materials

Publication details

Journal / source
ACS Applied Materials & Interfaces
Volume
18
Issue
2
Pages
4004–4014
Publication date
6 January 2026
Language