Wafer-Scale Fabrication of Miniaturized Graphene-Based Ion-Selective Field-Effect Transistor Chips for Clinical Plasma Test
Abstract
Rapid and reliable point-of-care plasma electrolyte analysis is critical for patient management under extensive care, driving demand for miniaturized sensors capable of ultralow sample consumption and multiplexed detection. We present a wafer-scale fabricated ion-selective graphene field-effect transistor (ISGFET) array that overcomes limitations of conventional ion-selective electrodes through a 2 μm × 5 μm sensing area and monolithic integration of K + /Na + /Ca 2+ selective membranes. The ISGFET sensor exhibited high sensitivity (52.43 mV/decade for K +, 53.62 mV/decade for Na +, and 28.72 mV/decade for Ca 2+ ) and selectivity in ion detection, as well as notable stability under physiological conditions and long-term reusability for over a month. The ISGFET-based sensing platform for the first time demonstrated its 100% diagnostic accuracy for potassium measurement in human plasma with an error of 4.37% (0.17 mM), when consuming less (
Fields of Study & Subjects
Analytical Chemistry and Sensors
Graphene research and applications
Advanced Sensor and Energy Harvesting Materials
Publication details
- Journal / source
- ACS Applied Materials & Interfaces
- Volume
- 18
- Issue
- 2
- Pages
- 4004–4014
- Publication date
- 6 January 2026
- Language
- EN