Extending Calibration Anchors Resolves Low- m / z Shifts in Negative-Mode Orbitrap Workflows

View at publisher
Authors: Francisco José Diaz Galiano, Bruno Le Bizec
Cited by: 1
References: 6

Abstract

Reliable mass accuracy is essential for the confident identification of diagnostic fragment ions in high-resolution mass spectrometry. During the analysis of perfluoroalkyl sulfonic acids (PFSAs) with a Q Exactive Orbitrap, we consistently observed the characteristic O 3 Ṡ – fragment, expected at a mass-to-charge ratio ( m / z ) of 79.9574, reported at m / z 79.9568, a systematic deviation of approximately −6 parts per million (ppm). Similar errors affected other ions below m / z 100, while precursors and higher- m / z fragments remained within the specification. Comparison with an Exploris 120 instrument, calibrated with anchors down to m / z 59, confirmed that the deviation is not intrinsic to the ion but originates from the limited calibration range of the Q Exactive standard calibration solution. We demonstrate that extending calibration to include additional low- m / z anchors generated in situ by in-source fragmentation fully corrects the error without affecting the accuracy at a higher m / z . This adjustment resolves systematic deviations for ions below m / z 100 in the Q Exactive instruments.

Fields of Study & Subjects

Per- and polyfluoroalkyl substances research Mass Spectrometry Techniques and Applications Molecular Sensors and Ion Detection

Publication details

Journal / source
Journal of the American Society for Mass Spectrometry
Publisher
American Chemical Society (ACS)
Volume
37
Issue
3
Pages
798–802
Publication date
3 February 2026
Language
EN