Matrix-Tolerant Quantification of THC and THCA in Complex Cannabis Products Using In-Sample Calibration with Multiple Isotopologue Reaction Monitoring
Abstract
High Resolution Image Download MS PowerPoint Slide Accurate quantification of Δ 9 -tetrahydrocannabinol (THC) and Δ 9 -tetrahydrocannabinolic acid (THCA) across diverse cannabis- and hemp-derived products remains challenging due to severe matrix effects, wide concentration variability, and the need for matrix-matched calibration in traditional LC–MS workflows. Here, we develop an in-sample calibration curve (ISCC) method based on multiple isotopologue reaction monitoring (MIRM) from stable-isotope-labeled (SIL) analytes to enable robust quantification of THC and THCA without external calibration curves. The approach leverages the theoretical relative isotopic abundances of SIL calibrators to generate multiple internal calibration points within each injection. By incorporating two SIL calibrators for THC (THC-D 3 and THC-D 9 ), applying a response-correction factor to harmonize labeled and unlabeled analytes, and utilizing native-analyte isotopologue transitions at high abundance, the method achieves a >600-fold dynamic range. The ISCC method demonstrated excellent linearity ( R 2 > 0.999), precision (
Fields of Study & Subjects
Forensic Toxicology and Drug Analysis
Isotope Analysis in Ecology
Cannabis and Cannabinoid Research
Publication details
- Journal / source
- Journal of the American Society for Mass Spectrometry
- Publisher
- American Chemical Society (ACS)
- Volume
- 37
- Issue
- 2
- Pages
- 548–555
- Publication date
- 26 January 2026
- Language
- EN