A flexible wireless skin patch for synchronized glucose monitoring and regulation
Abstract
Conventional diabetes management requires frequent invasive procedures such as finger-prick blood sampling and subcutaneous injections to coordinate glucose monitoring and medication. Here, we propose a novel, flexible, wearable, battery-free skin patch that synchronizes painless glucose monitoring and regulation capabilities with smartphone-mediated wireless control. This patch integrates bendable fluorescent hydrogel microneedles for minimally invasive glucose monitoring (50 to 450 mg/dL range) and thermoresponsive microneedles for metformin delivery. In diabetic mouse models, it accurately tracked interstitial glucose levels and, upon hyperglycemia detection, reduced blood glucose within 1 h (effects lasting 5-6 h). This system provides glucose monitoring with wireless data transmission and precise drug administration while eliminating pain, infection risk, and high costs. Its lightweight, disposable design offers a practical solution for improved diabetes care.
Fields of Study & Subjects
Advancements in Transdermal Drug Delivery
Advanced Sensor and Energy Harvesting Materials
Spectroscopy Techniques in Biomedical and Chemical Research
Publication details
- Journal / source
- Microsystems & Nanoengineering
- Publisher
- Springer Science and Business Media LLC
- Volume
- 12
- Issue
- 1
- Pages
- 23–23
- Publication date
- 8 January 2026
- Language