Continuous sensing technologies are expected to change how dynamic bioprocesses will be monitored and controlled in the future. However, a fundamental challenge in the field of biomolecular sensing is that low concentrations invariably cause slow sensor responses. In this paper, we explain how continuous sensors can be developed for quantifying very low biomarker concentrations with fast sensor response times. The measurement concept is based on combining reversible affinity-based nanoswitches with single-molecule readout. A generalizable rate-based simulation model was developed and validated on experimental data of a sandwich-based nanoswitch. The results show how the nanoswitch design and the sensing acquisition parameters control the amplitude and stochastic variations of the sensor response, and how these affect the precision of the concentration determination. We show that the measurement precision is limited by the counting statistics of molecular sandwich events. Using realistic design parameters, we predict limits-of-quantification in the low picomolar range within measurement timescales of minutes. These results pave the way for the development of intrinsically reversible nanoswitch sensors that can access unexplored concentration-time spaces of dynamic biosystems.

Fields of Study & Subjects

Advanced biosensing and bioanalysis techniques Advanced Biosensing Techniques and Applications Analytical Chemistry and Sensors

Publication details

Journal / source
Nature Communications
Publisher
Springer Science and Business Media LLC
Volume
17
Issue
1
Publication date
18 April 2026
Language
EN