Direct method for Ra-226 analysis in water samples using ICP-QQQ-MS

Read full text
Authors: Ivana Coha, Marko Štrok, Norbert Kávási
Cited by: 1
References: 48

Abstract

Abstract The paper describes a method of determination of Ra-226 in natural water using ICP-QQQ-MS, Agilent 8900, that enables an efficient direct determination of Ra-226 with high sensitivity without using any reaction gas and separation procedures for interference removal. The matrix effect, together with the instability of the sensitivity are controlled by spiking the sample with Ra-226. The method is simple and fast with a detection limit around of 0.1 fg g -1 (3.6 mBq kg -1 ), which is tenfold lower than required by the EU Directive for water consumption (40 mBq kg -1 ) and five times better than results obtained from alternative direct method for routine analysis described in the literature. The method can easily be implemented for routine analysis Ra-226 which was demonstrated by direct determination of Ra-226 in IAEA proficiency test water sample. This is the first time that the accuracy (trueness and precision) of a mass spectrometry method for the direct determination of Ra-226 has been tested and confirmed in an independent proficiency test.

Fields of Study & Subjects

Analytical chemistry methods development Radioactive contamination and transfer Pesticide Residue Analysis and Safety

Publication details

Journal / source
Scientific Reports
Publisher
Springer Science and Business Media LLC
Volume
16
Issue
1
Pages
3068–3068
Publication date
22 January 2026
Language