Evaluation of the visual field test algorithm: the Ambient Interactive ZEST-EX (AIZE-EX)

Read full text
Authors: Hiroki Nomoto, Chota Matsumoto, Keiji Yoshikawa, Mami Nanno, Shinji Kimura, Satoshi Inoue, Kenzo Yamanaka, Shunji Kusaka
Cited by: 2
References: 28

Abstract

Ambient Interactive ZEST EX (AIZE-EX) is a newly developed, time-saving visual field (VF) test algorithm for follow-up assessments in glaucoma patients that utilizes prior AIZE results. This study compared VF test results between AIZE and AIZE-EX in patients with glaucoma. Ninety-two eyes from 46 glaucoma patients were examined. All underwent the AIZE test at baseline and both AIZE and AIZE-EX tests at follow-up. Two test patterns were employed: 24-2 (44 eyes) and 24plus(1-2) (48 eyes). Mean deviation (MD), pattern standard deviation (PSD), mean sensitivity (MS), and test duration were compared between visits using the Mann-Whitney U test, and pointwise sensitivity differences were calculated. MD, PSD, and MS values obtained from AIZE-EX showed no significant differences compared with those from baseline and follow-up AIZE tests. Test duration for AIZE remained consistent between visits, whereas AIZE-EX reduced test time by approximately 10-30% (p

Fields of Study & Subjects

Ophthalmology and Visual Impairment Studies Glaucoma and retinal disorders Visual perception and processing mechanisms

Publication details

Journal / source
Scientific Reports
Publisher
Springer Science and Business Media LLC
Volume
16
Issue
1
Pages
5215–5215
Publication date
14 January 2026
Language