Differential sensitivity of impedance plethysmography and photoplethysmography sensors to temperature-induced peripheral vasoconstriction

Read full text
Authors: Seobin Jung, Seamus Thomson, Alexandros Pantelopoulos, Lindsey Sunden, Peter G. Richards, Shwetak Patel, Sam Sheng
Cited by: 1
References: 20

Abstract

Abstract Impedance plethysmography (IPG) and photoplethysmography (PPG) are non-invasive techniques for measuring blood volume changes. This study investigated the differential responses of IPG and PPG to temperature-mediated vasoconstriction induced by localized cooling. Twenty-one participants underwent control and treatment conditions, with fake or real ice cubes applied to the forearm. Blood pressure remained stable, while heart rate decreased. PPG signal amplitude significantly decreased with cooling (p adj = 0.004), indicating sensitivity to superficial blood flow changes. In contrast, IPG signal amplitude remained stable (p adj = 1.0). No statistically significant differences were observed in timing-derived metrics. These findings suggest IPG is less sensitive to superficial changes in blood flow than PPG, and may be more suitable for monitoring deeper blood flow. This study provides insights into the distinct sensitivities of IPG and PPG, with implications for wearable device development and cardiovascular monitoring.

Fields of Study & Subjects

Non-Invasive Vital Sign Monitoring Hemodynamic Monitoring and Therapy Electrical and Bioimpedance Tomography

Publication details

Journal / source
Scientific Reports
Publisher
Springer Science and Business Media LLC
Volume
16
Issue
1
Pages
6828–6828
Publication date
31 January 2026
Language