Randomized trial of smartphone application and bed sensor for atrial fibrillation detection in high-risk patients

Read full text
Authors: Joonas Lehto, Joel Nuotio, Arto Relander, Jussi Jaakkola, Olli Lahdenoja, Tuija Vasankari, Arman Anzanpour, Ismail Elnaggar, R. Rekola, Jonas Sandelin, Tero Hurnanen, Juhani K.E. Airaksinen
+2 more Tero Koivisto, Tuomas Kiviniemi
Cited by: 0
References: 29

Abstract

This two-arm single-center exploratory randomized controlled trial evaluated the efficacy of prolonged rhythm monitoring in atrial fibrillation (AF) detection after an invasive cardiac procedure. Altogether 150 patients were enrolled. In the intervention group (IG), a bed sensor (EMFIT QS) and twice-daily smartphone recordings (CardioSignal app) were used, followed by a 12-lead ECG and a continuous three-to-seven-day ECG monitoring if alerts occurred. The control group (CG) received usual care. Overall, 78 patients were assigned to the IG and 72 to CG. During the three-month follow-up, AF was detected in 6/78 (7.7%) patients in the IG and in 0/72 (0.0%) in the CG (absolute risk difference 7.7%, 95% CI 1.8–13.6%, p = 0.029). After exclusion of patients who withdrew before the 3-month follow-up, 33/68 (48.5%) patients had alarms not leading to ECG-verified AF diagnosis, indicating that the current approach, in its present form, is not suitable for routine clinical implementation. Future studies should concentrate on minimizing alarms not leading to AF diagnosis when developing these novel non-ECG-based technologies. ClinicalTrials.gov Identifier: NCT05351775, 2022/04/28.

Fields of Study & Subjects

Atrial Fibrillation Management and Outcomes ECG Monitoring and Analysis Non-Invasive Vital Sign Monitoring

Publication details

Journal / source
Scientific Reports
Volume
16
Issue
1
Pages
7088–7088
Publication date
3 February 2026
Language
EN