Optimizing the dissection of small-diameter pulmonary vessels using vessel-sealing systems

Read full text
Authors: Yuichiro Ueda, Jun‐ichi Wakahara, So Miyahara, Hiroyasu Nakashima, Yoshiko Masuda, Fumihiro Shoji, Takeshi Shiraishi, Toshihiko Sato
Cited by: 0
References: 14

Abstract

Although vessel-sealing systems have been shown useful for dissection of small-diameter vessels, no studies to date have evaluated the optimal angle of dissection. Lateral thermal damage to a canine pulmonary artery was assessed microscopically. Burst pressure was compared in swine carotid arteries undergoing orthogonal and oblique dissections to determine the safer dissection technique. Histological analysis of a canine model revealed a thermal spread of approximately 1.3 mm. Burst pressure comparison showed no significant differences between orthogonal and oblique dissection angles (829.2 mmHg [range 608-1214 mmHg] vs. 949.1 mmHg [range, 593-1306 mmHg], P = 0.206). In a surrogate arterial model, burst pressure did not differ between orthogonal and oblique cuts. Given histologically limited lateral thermal spread on PA, angle may be less critical than avoiding vessel tension. These findings may contribute to the safer application of vessel-sealing systems in patients undergoing minimally invasive thoracic surgery.

Fields of Study & Subjects

Vascular Procedures and Complications Hemostasis and retained surgical items Surgical Sutures and Adhesives

Publication details

Journal / source
Scientific Reports
Volume
16
Issue
1
Publication date
11 February 2026
Language