Development and preclinical results of a transcatheter aortic valve implant with novel polymeric leaflets
Abstract
Current transcatheter aortic valve implantation (TAVI) systems use glutaraldehyde-fixed tissue leaflets, which are prone to structural valve degeneration. Polymeric valves may improve durability, hemocompatibility, and manufacturability. We designed, produced, and evaluated a self-expanding nitinol-framed TAVI device with siloxane poly(urethane-urea) (SiPUU) leaflets and an electrospun SiPUU skirt. Bench testing included frame fatigue to 200 million cycles, hydrodynamics per ISO 5840-3, and flow visualization using particle image velocimetry. Biocompatibility was assessed per ISO 10993. Preclinical evaluation involved nine ovine implants with serial imaging and histology to 90 days. In vitro tests showed EOAs of 1.3-1.5 cm² and 1.7-2.1 cm² (21 and 24 mm annulus, respectively, simulating under expansion), regurgitant fraction
Fields of Study & Subjects
Cardiac Valve Diseases and Treatments
Electrospun Nanofibers in Biomedical Applications
Aortic Disease and Treatment Approaches
Publication details
- Journal / source
- npj Cardiovascular Health
- Publisher
- Springer Science and Business Media LLC
- Volume
- 3
- Issue
- 1
- Publication date
- 23 March 2026
- Language
- EN