A micro-organ based microfluidic biosensor for continuous monitoring of glucose levels in vivo

Read full text
Authors: Emilie Puginier, Antoine Pirog, F. Macouillard Poulletier de Gannes, Julien Gaitan, Annabelle Hurtier, Dorian Chapeau, Marie Monchablon, Matthieu Raoux, S. Renaud, J Lang
Cited by: 2
References: 61

Abstract

Abstract Current approaches to detect analytes use biomolecules or single cells and thus do not harness the computational power of endogenous algorithms present in micro-organs. Consequently, current on-line detection of glucose does not allow an autonomous artificial pancreas. In contrast, monitoring a few electrogenic pancreatic islets may provide a more appropriate read-out. Indeed, upon stimulation islets react with so-called slow potentials (SP) and their amplitude reflects the degree of intercellular coupling. We have now developed a microfluidic microelectrode chip containing a few islets and linked to interstitial fluids in live rats by microdialysis. Blood and dialysate glucose were determined concomitantly with ex-vivo islet electrical activity during an intraperitoneal glucose or insulin challenge. Blood glucose was tightly correlated to islet SP frequency, and to a lesser degree to SP amplitudes. This demonstrates the feasibility and usefulness of microorgan-based biosensors which may also be of interest in the therapy of diabetes.

Fields of Study & Subjects

Pancreatic function and diabetes Microfluidic and Bio-sensing Technologies 3D Printing in Biomedical Research

Publication details

Journal / source
npj Biosensing
Publisher
Springer Science and Business Media LLC
Volume
3
Issue
1
Publication date
19 February 2026
Language
EN