Point-of-care multiplexed detection of cytokines using a HexaPie electrode array for asthma endotyping

Read full text
Authors: Nathan Kodjo Mintah Churcher, Fatima Zia Rizvi, Aleeza Qureshi, Salsabil Aziz, Sasya Madhurantakam, Nana Kwame Ayisi‐Boateng, Ernest Adankwah, Michael Kofi Ansah, Richard Odame Phillips, Shalini Prasad
Cited by: 1
References: 30

Abstract

Abstract Accurate multiplexed cytokine quantification in complex biofluids is essential for immune monitoring and respiratory disease phenotyping. We report HexaPie, a label-free, non-faradaic impedance biosensor that enables simultaneous detection of IL-8, IL-10, and IP-10 in human saliva through electrical double-layer (EDL) modulation at spatially multiplexed gold microelectrodes. The platform demonstrated limits of detection (LOD) of 1.17 ng mL −1 (IL-8), 12.5 pg mL −1 (IL-10), and 0.20 ng mL −1 (IP-10) with coefficients of variation below 15% and spike–recovery rates within 80–120%, satisfying CLSI analytical performance criteria. Specificity studies confirmed minimal signal interference from bovine serum albumin, uric acid, and ascorbic acid, validating robust performance in native saliva. Comparative clinical evaluation using ELLA and Luminex xMAP immunoassays revealed strong Bland–Altman concordance for IL-8 and IP-10 (mean bias

Fields of Study & Subjects

Advanced Biosensing Techniques and Applications Biosensors and Analytical Detection Microfluidic and Bio-sensing Technologies

Publication details

Journal / source
npj Biosensing
Publisher
Springer Science and Business Media LLC
Volume
3
Issue
1
Publication date
2 March 2026