Initial testing procedure based on microextraction followed by LC-HRMS analysis to determine 107 xenobiotics and their metabolites in serum/plasma

View at publisher
Authors: Monica Mazzarino, Koen Deventer, Laurie De Wilde, K. Roels, Peter Van Eenoo
Cited by: 2
References: 17

Abstract

) carry-over (no signals in the negative sample injected after the positive sample at high concentration), matrix effect (10-28%), extraction yield (42-89%), and extract stability (the analytes were stable for at least 72 h in the autosampler at 10 °C). The method was successfully applied to the analysis of samples containing the compounds at low nanogram per milliliter range, demonstrating its effectiveness for doping control purposes. Stability studies showed that the compounds were stable for at least 3 months at -20 and 4 °C in serum and plasma samples. In contrast, at 22 °C several thiazide-based compounds were completely degraded after 4 weeks; FG2216 was no longer detectable after 7 weeks; S6 and RAD140 were no longer detectable after 9 weeks, whereas trenbolone was completely degraded after 14 weeks. The other compounds were still visible for the entire study period, with variations in the range of 37-56%.

Fields of Study & Subjects

Analytical chemistry methods development Pesticide Residue Analysis and Safety Carcinogens and Genotoxicity Assessment

Publication details

Journal / source
The Analyst
Publisher
Royal Society of Chemistry (RSC)
Volume
151
Issue
5
Pages
1477–1494
Publication date
1 January 2026
Language
EN