A candidate reference measurement procedure for aldosterone measurement in human serum and urine on the basis of isotope dilution liquid chromatography-tandem mass spectrometry

View at publisher
Authors: Siyu Chen, Junyi Wu, Zihan Liu, Yuchen Wan, Chuanfen Xie, Yingguo Wang, Zechen Liu, Ke Chen, Jiashu Yang, Xinyi Rao, Hui Yuan
Cited by: 0
References: 30

Abstract

. The CV ranged from 1.1% to 1.4%. The recoveries of both samples were within the range of 95-105%. This method meets the analytical requirements of a reference measurement procedure, reduces the use of hazardous chemicals during sample preparation, and may serve as a candidate reference method for the accurate quantification of aldosterone in serum and urine samples in the laboratory.

Fields of Study & Subjects

Hormonal Regulation and Hypertension Analytical chemistry methods development Scientific Measurement and Uncertainty Evaluation

Publication details

Journal / source
Analytical Methods
Volume
18
Issue
7
Pages
1492–1502
Publication date
1 January 2026
Language