Expanding luminol activity: UV-enhanced peroxidase-like activity for dual-mode wide-pH biosensing

Read full text
Authors: Mohamed Hassan Mahana, Syed Rahin Ahmed, Poushali Das, Seshasai Srinivasan, Amin Reza Rajabzadeh
Cited by: 2
References: 62

Abstract

and dopamine with limit of detection (LOD) values of 0.0149 mM and 0.52 µM, respectively. This study also reveals that UV activation substantially enhances luminol's peroxidase-like activity, decreasing the detection time from 35 to 4.5 min and improving the sensitivity by lowering the LOD from 0.0149 to 0.0085 mM. Kinetic studies using Michaelis-Menten and Lineweaver-Burk analyses demonstrate that luminol exhibits high peroxidase-like catalytic competency compared with other reported nanozymes. Moreover, the method was successfully applied for the determination of dopamine in a simulated blood sample, achieving an LOD value of 1.5089 µM, thereby demonstrating its excellent performance and suitability for dopamine analysis in complex biological environments. These findings highlight the use of luminol as an efficient, UV-enhanced peroxidase mimic with strong potential as a rapid and sensitive platform for next-generation colorimetric biosensors in clinical applications.

Fields of Study & Subjects

Advanced Nanomaterials in Catalysis Nanoplatforms for cancer theranostics Neutrophil Myeloperoxidase and Oxidative Mechanisms

Publication details

Journal / source
Analytical Methods
Publisher
Royal Society of Chemistry (RSC)
Volume
18
Issue
11
Pages
2225–2238
Publication date
1 January 2026
Language