A two-point fluorescence detection flow cell with empirical calibration for high optical density analysis

Read full text
Authors: Tin Weitner, Davor Šakić
Cited by: 1
References: 15

Abstract

A two-point fluorescence detection flow cell with empirical calibration is presented for correcting inner-filter effects at high optical densities. Using only fluorescence signals from fixed detection points, linear fluorescence-concentration relationships are restored under static sample and flow conditions, demonstrating a general strategy for quantitative measurements in compact flow systems.

Fields of Study & Subjects

Optical Imaging and Spectroscopy Techniques Innovative Microfluidic and Catalytic Techniques Innovation Water Quality Monitoring and Analysis

Publication details

Journal / source
The Analyst
Publisher
Royal Society of Chemistry (RSC)
Volume
151
Issue
10
Pages
2767–2772
Publication date
1 January 2026
Language