Foreign Body Reaction in Wrist Arthroplasty Using a Carbon Fiber Reinforced Poly-Ether-Ether-Ketone Articulation: A Secondary Analysis of 11 Cases

Read full text
Authors: Daniel Reiser, ARPAD SZALLASI, Severin Sape, Per F. Wretenberg, Marcus Sagerfors
Cited by: 0
References: 32

Abstract

Abstract Wrist arthroplasty is a motion-preserving procedure. However, despite improvements in implant design, implant failure remains an issue. The aim of this secondary analysis was to assess failures associated with synovitis and to report histopathological findings in an initial patient cohort (n = 40) reported previously. Eleven cases of implant failure and synovitis were identified among hemi and total wrist arthroplasties with a carbon fiber reinforced poly-ether-ether-ketone (CFR-PEEK) articulation. Six cases had undergone component exchange/revision; of these, five were revised to a radiocarpal arthrodesis. Pronounced wear of the CFR-PEEK articulating surface was noted. Histopathological examination indicated that CFR-PEEK particles were incorporated by macrophages, but no giant cell reactions could be seen. A high frequency of synovitis warranting reoperation was noted. We recommend caution and close monitoring of CFR-PEEK articulations in the wrist.

Fields of Study & Subjects

Orthopedic Surgery and Rehabilitation Orthopedic Infections and Treatments Musculoskeletal synovial abnormalities and treatments

Publication details

Journal / source
Journal of Wrist Surgery
Publication date
7 January 2026
Language
EN