High performance two-dimensional photon-counting imaging detector based on delay line anode

View at publisher
Authors: Yalong Zhang, Yongan Liu, Xianghui Yang, Zhe Liu, Lizhi Sheng
Cited by: 0
References: 17

Abstract

A microchannel plate detector based on a direct-collection cross-delay-line (XDL) anode readout is proposed. By converting spatial information into temporal signals and processing them with a high-speed time-to-digital converter, the system achieves a balance between high spatial resolution and high counting rate. The article elaborates on the detector's design principles, image encoding system, and signal processing workflow. Experimental evaluations were conducted on key performance metrics, including the XDL encoding system's nonlinearity, timing accuracy, as well as the overall spatial resolution and maximum counting rate of the detector. Test results demonstrate that the detector exhibits a spatial resolution better than 70 μm, a maximum counting rate of 1 MHz, and excellent timing precision (single-channel accuracy

Fields of Study & Subjects

Photocathodes and Microchannel Plates Particle Detector Development and Performance Radiation Detection and Scintillator Technologies

Publication details

Journal / source
Review of Scientific Instruments
Volume
97
Issue
3
Publication date
1 March 2026
Language
EN