Abstract

Scanning nitrogen-vacancy (NV) center electrometry has shown potential for quantitative quantum imaging of electric fields at the nanoscale. However, achieving nanoscale spatial resolution remains a challenge since employing gradiometry to overcome electrostatic screening causes resolution-limiting trade-offs including the averaging effect and the sensor-sample proximity. Here, we demonstrate a scanning NV center protocol that achieves an enhanced spatial resolution of ∼10 nm. We develop an axially symmetric probe with a sub-nanometer oscillating amplitude, which simultaneously provides robust intermittent-contact mode feedback and ensures close engagement between the diamond tip and the sample. As an example, we experimentally demonstrate a 10 nm spatial resolution on ferroelectric lithium niobate. Scanning NV center electrometry with this resolution can directly resolve the nanoscale polar textures and dynamics of emerging ferroelectrics, which commonly arise on the scale of tens of nanometers.

Fields of Study & Subjects

Nuclear Physics and Applications Laser-induced spectroscopy and plasma Integrated Circuits and Semiconductor Failure Analysis

Publication details

Journal / source
Review of Scientific Instruments
Volume
97
Issue
4
Publication date
1 April 2026