BACKGROUND: Electrode tip fold-over is a critical complication in cochlear implantation (CI), potentially compromising hearing outcomes. Reliance on postoperative X-ray for confirmation delays detection and exposes patients to radiation. OBJECTIVES: This study assessed intraoperative transimpedance matrix (TIM) for detecting CI632 slim modiolar electrode (SME) tip fold-over, comparing its performance with postoperative X-ray. MATERIAL AND METHODS: = 26) CI632 electrode. Immediately after electrode insertion, intraoperative TIM measurements were performed using the Custom Sound EP software, generating heat maps to visualize the intracochlear voltage distribution. Postoperative X-ray imaging was performed to confirm electrode positioning. RESULTS: Electrode tip fold-over occurred in 7.69% of cases (2/26). TIM heatmaps provide real-time visualization of electrode positioning and enable immediate intraoperative correction. In cases where X-ray imaging suggested residual post-revision fold-over, subsequent computed tomography (CT) scans confirmed proper electrode positioning, which highlights the potential of TIM to reduce false positives. CONCLUSION AND SIGNIFICANCE: TIM is a highly effective, noninvasive tool for real-time intraoperative monitoring of CI electrode positioning, with superior accuracy to X-ray imaging without radiation exposure. Integration of TIM into the surgical workflow can reduce the need for revision surgery and improve patient outcomes.

Fields of Study & Subjects

Hearing Loss and Rehabilitation Ear Surgery and Otitis Media Hearing Cochlea Tinnitus Genetics

Publication details

Journal / source
Acta Oto-Laryngologica
Volume
146
Issue
4
Pages
416–428
Publication date
28 March 2026
Language
EN