Thorough biomechanical analysis of arterial response to EasyEndo-Lite staple rotation: a simulation study in abaqus

View at publisher
Authors: Azizeh Hosseinjany, Pinar Çağan, Ali Kimiaei, Seyedehtina Safaei, Demircan Canadinç, Kutlu Ca
Cited by: 0
References: 33

Abstract

Understanding instrument-tissue interaction is vital for safe surgery. We used Abaqus to simulate arterial response to EasyEndo-Lite Staple rotation, focusing on abrupt motion and strain-rate effects. Sudden angular rotation produced high strain rates and peak stresses in the clamped arterial segment, reaching 1.4 MPa at 15° and ∼3.7 MPa at 30°. Rapid cessation of rotation increased forces by up to 28%, indicating elevated damage risk. The simulations provided detailed deformation and stress data across scenarios, demonstrating how optimizing instrument design and operating parameters could substantially reduce tissue trauma and improve surgical outcomes.

Fields of Study & Subjects

Elasticity and Material Modeling Soft Robotics and Applications Surgical Sutures and Adhesives

Publication details

Journal / source
Computer Methods in Biomechanics & Biomedical Engineering
Pages
1–12
Publication date
16 February 2026
Language
EN