Prompt gamma activation analysis - neutron imaging instrument on the China Mianyang research reactor

View at publisher
Authors: Yuhua Ma, Xi Yang, Hang Li, Hongwen Huang, Yang Wu, Heyong Huo, Wei Yin, Sheng Nan Wang, Bin Liu, Rundong Li, Ping Wang
Cited by: 0
References: 32

Abstract

To meet the application demand for simultaneous structure inspection and elemental analysis, research on neutron imaging-driven prompt gamma activation analysis instrument is of great significance. Relying on the China Mianyang Research Reactor, the China Academy of Engineering Physics proposed the physical design of the PGAA -NT apparatus, carried out Monte Carlo simulations, signal commissioning and experimental tests, and successfully established an experimental facility that couples PGAA with neutron imaging. In terms of characterisation, the beam parameters, background features, Compton-suppression performance, wide-energy-range detection efficiency, and imaging spatial resolution of the instrument were accurately measured. The instrument exhibits the low background (full spectrum

Fields of Study & Subjects

Nuclear Physics and Applications Radiation Detection and Scintillator Technologies Nuclear reactor physics and engineering

Publication details

Journal / source
Nondestructive Testing And Evaluation
Pages
1–20
Publication date
11 January 2026
Language
EN