Measurement of flame temperature using active and passive tunable diode laser absorption spectroscopy (TDLAS)

View at publisher
Authors: Xiaodong Huang, Zhiling Li, Xiaoyong Zhang, Yanan Bao
Cited by: 0
References: 28

Abstract

Reconstruction accuracy is a crucial indicator for assessing the combustion conditions in combustion diagnostics. Existing flame-temperature measurement methods require comparison and verification using simulations, infrared thermal imaging cameras, and thermocouples, which increase measurement complexity and reduce efficiency. This study introduces a method based on tunable diode laser absorption spectroscopy (TDLAS) that utilizes an array sensor to capture specific absorption and radiation spectral signals to measure the combustion field temperature. Active and passive spectral information at different temperatures was obtained through calibration experiments. The temperature field of the K+-doped premixed stable flame was imaged. The difference between the active and passive spectral temperature field images did not exceed 8%, and the maximum relative error compared to the thermocouple measurements did not exceed 6%, demonstrating good reconstruction accuracy. Simultaneous active and passive spectral temperature measurements provide a novel approach for developing high-precision flame measurements and reducing system complexity.

Fields of Study & Subjects

Spectroscopy and Laser Applications Combustion and flame dynamics Laser-induced spectroscopy and plasma

Publication details

Journal / source
Instrumentation Science & Technology
Volume
54
Issue
5
Pages
620–634
Publication date
6 January 2026