Finite element analysis of hydroxyapatite coatings in minimizing stress and displacement in middle ear prostheses
Summary
No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.
Summarise with VIRAT Free to try with your Science Catalogs account.Abstract
This study examines the biomechanical performance of a middle ear prosthesis and evaluates the effect of a hydroxyapatite (HA) coating using finite element (FE) analysis under a 90 dB sound pressure level within the 250–8000 Hz range. Prosthesis materials made of titanium grade 2, PEEK, and bone cement were assessed. Bone cement exhibited the highest stress peaks (≈1.27 × 10−8 MPa) with pronounced oscillations along the Z-distance, while PEEK presented medium stress levels (≈3.78 × 10−9 MPa). Titanium grade 2 demonstrated the lowest stress response (
Fields of Study & Subjects
Ear Surgery and Otitis Media
Orthopaedic implants and arthroplasty
Bone Tissue Engineering Materials
Publication details
- Journal / source
- Mechanics of Advanced Materials and Structures
- Volume
- 33
- Issue
- 1
- Publication date
- 30 March 2026
- Language
- EN