Finite element analysis of hydroxyapatite coatings in minimizing stress and displacement in middle ear prostheses

View at publisher Summarise with VIRAT
Authors: Naili Taki Eddine, Ait Kaci Djafar, Lahcene Ghouari, A. Sahli, Khatir Omar, Abdelmadjid Moulgada, Benatmane Charaf eddine, Murat Yaylacı
Cited by: 0
References: 63

Summary

No summary yet. VIRAT can read this paper and write a plain-language summary of its aim, methods, key findings and limitations.

Summarise with VIRAT Free to try with your Science Catalogs account.

Abstract

This study examines the biomechanical performance of a middle ear prosthesis and evaluates the effect of a hydroxyapatite (HA) coating using finite element (FE) analysis under a 90 dB sound pressure level within the 250–8000 Hz range. Prosthesis materials made of titanium grade 2, PEEK, and bone cement were assessed. Bone cement exhibited the highest stress peaks (≈1.27 × 10−8 MPa) with pronounced oscillations along the Z-distance, while PEEK presented medium stress levels (≈3.78 × 10−9 MPa). Titanium grade 2 demonstrated the lowest stress response (

Fields of Study & Subjects

Ear Surgery and Otitis Media Orthopaedic implants and arthroplasty Bone Tissue Engineering Materials

Publication details

Journal / source
Mechanics of Advanced Materials and Structures
Volume
33
Issue
1
Publication date
30 March 2026
Language
EN